Pulmodyne VT Select Manual Resuscitator

by Intersurgical

Description

Pulmodyne's new VT Select Manual Resuscitator with unique timed release valve, provides tactile feedback with the goal of eliminating hyperventilation. The timed release valve restricts the refill of the bag to approximately 4.5 seconds and is intended to help the user deliver the AHA recommended 10 breaths per minute rate.

 

 

  • Bag Volume: 1200mL.
  • Timed release valve can be switched off allowing for more traditional bag refill time if desired
  • Low profile strap and finger ridges help ensure secure grip
  • Optional accessories include PEEP valve, Exhalation Filter and Manometer
Item #
Description
In Stock
Price
Quantity
 
# N/A
View alternatives

Accessories

Item #
Description
In Stock
Price
Quantity
 
# N/A
View alternatives

More Information

 

Pulmodyne VT Select Video with Forrest
The VT Select resolves issues of bagging too fast & delivering too much volume when ventilating a patient. See the video to discover more!
Pulmodyne VT Select Video
See the new VT Select Manual Resuscitator in action in this video.

Questions and Answers

Q

How much volume does the VT Select provide when a breath is delivered?

Anonymous- On Friday, 24 June 2022

A

When using the VT Select BVM, the delivered volume range for a 1 handed technique is 530mL and 690mL for a 2-handed technique.

Sara K.- answered On Friday, 24 June 2022

Q

Can I use the VT Select without using the hand strap?

Anonymous- On Friday, 24 June 2022

A

The VT Select BVM hand strap is designed to be low profile, so a user can grip around the strap if desired.

Sara K.- answered On Friday, 24 June 2022

Q

How does the VT Select timed release valve get activated?

Anonymous- On Friday, 24 June 2022

A

The VT Select BVM comes packaged with the timed refill valve switched to the active mode. The user can begin using the bag immediately once opened.

Sara K.- answered On Friday, 24 June 2022

Similar Products